SPECIFICATIONS
Model | PR224 | PR224/E | PR423/E | PR4202 | PR4202/E |
Maximum Capacity (g) | 220 | 420 | 4200 | ||
Readability d (g) | 0.0001 | 0.001 | 0.01 | ||
Pan Size | 3.5 in (90 mm) | 4.7 in (120 mm) | 7.1 in (180 mm) | ||
Internal Calibration | InCal - Semi Automatic | Not applicable | |||
Communication | RS232 | ||||
Assembled Dimensions | 317 mm x 303 mm x 201 mm (LxHxW) | 317 mm x 93 mm x 201 mm (LxHxW) | |||
Repeatability (sd.), 5% if Full Load to Full Range (g) | 0.00008 | 0.0008 | 0.008 | ||
Linearity Deviation, Tryical (g) | ±0.00006 | ±0.0006 | ±0.006 | ||
Linearity ± (g) | 0.0002 g | 0.002 g | 0.02 g | ||
Stabilization Time (s) | 3g | 2g | 1g | ||
Display | LCD with backlight | ||||
Minimum Weight (typical) (g) (USP, K=2, U=0.10%) | 0.16 | 1.6 | 16 | ||
Min-Weight (Optimal) (g) (USP, K=2, U=0.10%, SRP=<0.41) | 0.082 | 0.82 | 8.2 | ||
Unit | Miligram, Gram, Kilogram, Ounce, Carat, Pennyweight, Ounce Troy, Newton, Grain, Pound, Taiwan Tael, Custom | Gram, Kilogram, Ounce, Carat, Pennyweight, Ounce Troy, Newton, Grain, Pound, Taiwan Tael, Custom | |||
Applications | Basic Weighing, Parts Counting, Percent Weighing | ||||
Tare Range | To full range by subtraction | ||||
Tare Time | 1 | ||||
Power Supply | Power Input: 100-240V - 200mA 50-60Hz 12-18VA Power Output: 12 VDC 0.5A | ||||
Operating Temperature Range | Operating conditions for ordinary lab application: +10 to 30°C (operability guaranteed between +5 and 40°C)) | ||||
Net Weihght | 4.5 kg | 3.5 kg |