The GENESYS 180 UV-Vis Spectrophotometer provides double beam operation for laboratories requiring a reference cell position for applications such as kinetics. An 8-cell carousel comes standard for high-throughput applications. The spectrophotometer includes room-light resistance, allowing lid-open operation for improved speed and convenience. A full line of accessories, including automated cell changers and cell holders, sipper, and fiber optic probes for measurments outside of the sample compartment are designed to simplify sampling and accommodate high throughput measurements and temperature control.
฿0
฿0
ราคาสมาชิก
 
฿0
ราคาสมาชิก
฿0

รายละเอียดสินค้า

SPECIFICATIONS

 Model GENESYS10
 Baseline Flatness ±0.002 A
 Compartments 8-position cell changer (standard)
 Also accommodates (optional):
 4-position cell changer (long path cells)
 Peltier thermostatted cell holder (20-60 °C)
 Sipper accessory
 Fiber optic probe coupler
 Connections Single USB-A
 Duplex USB-A
 Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor,  available
 Print via USB, Ethernet or Wi-Fi USB adaptor available
 Description GENESYS 180 UV-Vis Spectrophotometer
 Detector Type Dual Silicon Photodiodes
 Dimensions (L x W x H) 7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels
 Spectral Bandwidth 35.5 x 38.5 x 19.5 cm
 Wavelength Range 190 to 1100 nm
 Drift (At 500 nm after 1 hour warmup) <0.0005 A/hr.
 Electrical Requirements
 External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz
 Keypad Touchscreen
 Lamp Life > 5 years typical, 3 years guaranteed
 Noise (RMS at 500 nm 60 consecutive measurements)0.00020A at 0A at 260 and 500 nm
 0.00030A at 1A at 260 and 500 nm
 0.00040A at 2A at 260 and 500 nm
 Optical Design Double Beam
 Photometric Accuracy ±0.002A at 0.5A, ±0.004A at 1.0A, ±0.008 A at 2.0 A
 Photometric Display -3A to +5A
 Photometric Range 2A to +3.5A
 Photometric Repeatability ±0.001 A at 1 A measured at 1.0 A at 546 nm
 Photometric Display
 -3A to +5A
 Photometric Range 2A to +3.5A
 Photometric Repeatability ±0.001 A at 1 A measured at 1.0 A at 546 nm
 Printer Included Snap-on printer available
 Scan Speed Slow, medium and fast - up to 1600 nm/min.
 Stray Light
 < 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2)
 Warranty
 2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase
 Wavelength Accuracy
 ±0.5 nm
 Wavelength Data Interval
 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm
 Wavelength Repeatability < ±0.2 nm
 Weight (Metric) 7.5 kg
 Unit Size Each